SDS-PAGE gel trouble shooting
One of the favorite quotes in the lab is “the gel never lies”. This is a handy compiled list adapted from “gradigel” product literature.
TROUBLE SHOOTING
Problem: Distorted protein bands.
Cause: Air bubbles in the sample wells, or between gel and cassette, or at the bottom of the cassette.
Solution: Use a transfer pipette to displace the air bubbles from the sample wells.
Problem: Streaking.
Cause: Poorly soluble or weakly charged particles (such as carbohydrates) in sample.
Solution:
1. Centrifuge sample.
2. Change pH of buffer.
3. Heat sample in the presence of SDS.
Problem: Bands difficult to distinguish.
Cause: Incorrect gel selection, sample overloading, insufficient cooling buffer.
Solution:
1. Reduce sample size.
2. Select a gel which separates in the desired molecular weight range.
3. For proteins of similar molecular weight a longer gel may be useful.
4. Increase buffer in outer tank.
Problem: Sample spreading across the gel.
Cause: Too much salt in the sample.
Solution: Reduce salt by dialysis or ultra-filtration.
Problem: Sample contains appreciable carbohydrate.
Solution: Enzymatically remove the carbohydrate.
Problem: Sample contains lipoproteins.
Solution: Use a gel with large pore size at top. Try addition of a non-ionic detergent.
Problem: Protein denaturation and band inversion.
Cause: Excessive heating.
Solution: Always start with chilled buffer (<15°C).
Problem: Diffuse protein zones in gel after staining.
Cause: SDS still present in gel.
Solution: Fix in 10% TCA prior to staining, otherwise use 30% methanol in the destaining solution.
Comments
Leave a Reply
You must be logged in to post a comment.